EC0x Specifications (common to all EC0x chambers)

Controller Section : Model TC01 (included in chamber cost) (also available for RETROFIT)

Temperature Setpoint Range : -73 Deg C to +300 Deg C in 0.1 Deg C Steps (LN2 Option; -100 Deg C to +300 Deg C)

Time at Temperature Range : 0.1 to 1800 Minutes or Continuous

Temperature Ramping Rate Control : Not Controlled Locally

Programmable Set Temps and Times: 10 Programmable Set Temperatures and Times

Absolute Error (not including probe error) : +/- 0.4 Deg C

Long Term Stability (per month) : +/- 0.3 Deg C

Temperature Resolution : 0.1 Deg C

Line Voltage Sensitivity (105-125 VAC) : +/- 0.2 Deg C

Programmable Temperature Loop Control : 1 to 1800 Cycles of Programmable Set Temperatures and Set Times or Continuous

Temperature Control Technique : PID Algorithm / Pulse Width Modulated, Programmable from Remote Interfaces

Local Operation : 16 Key Keypad / 4 Digit Display

Remote Operation : Built-in IEEE-488 / RS232 Interfaces

Additional I/O Capability : IEEE-488 to RS232 Transparent Mode, 2 Auxiliary Outputs, 1 Auxiliary Input

Safety Features : Open/Short Probe Detection, Adjustable Over Temperature Thermostat, Software Upper Temperature Limit

Convenience Features : Heat and Cool Enable/Disable Switches

Home